KEY COMPARISON

Final report on key comparison EUROMET.EM-K4 of 10 pF and 100 pF capacitance standards

Published under licence by IOP Publishing Ltd
, , Citation J H Belliss 2002 Metrologia 39 01004 DOI 10.1088/0026-1394/39/1A/4

0026-1394/39/1A/01004

Abstract

This report describes a comparison of 10 pF and 100 pF capacitance standards organized within the framework of EUROMET Project No. 345 between 1995 and 1998. Piloted by the National Physical Laboratory (NPL, UK), the comparison involved sixteen national metrology institutes (NMIs), several of which are members of other regional metrology organizations, and the Bureau International des Poids et Mesures. The results show that there are significant differences between representations of the farad by some NMIs. However, the agreement demonstrated by the comparison gives confidence in maintaining traceability for the farad via either a calculable capacitor or a quantum Hall reference standard and the consensus value of the von Klitzing constant, RK-90.

Export citation and abstract BibTeX RIS

10.1088/0026-1394/39/1A/4