Brought to you by:
KEY COMPARISON

Activity measurements of the radionuclide 99mTc for the IFIN-HH, Romania in the ongoing comparison BIPM.RI(II)-K4.Tc-99m

, , , and

Published under licence by IOP Publishing Ltd
, , Citation C Michotte et al 2014 Metrologia 51 06014DOI 10.1088/0026-1394/51/1A/06014

0026-1394/51/1A/06014

Abstract

In 2013, a comparison of activity measurements of 99mTc using the Transfer Instrument of the International Reference System (SIRTI) took place at the Institutul National de Cercetare – Dezvoltare in Fizica si Inginerie Nucleara – 'Horia Hulubei' (IFIN-HH, Romania). Ampoules containing about 21 kBq and 40 kBq of a 99mTc solution were measured in the SIRTI for, respectively, one and three half-lives. The comparison, identifier BIPM.RI(II)-K4.Tc-99m, is linked to the BIPM.RI(II)-K1.Tc-99m comparison and the degrees of equivalence with the key comparison reference value and between the present IFIN-HH result, the other K4 participants and the six participants in the K1 comparison have been evaluated.

Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Export citation and abstractBibTeXRIS

10.1088/0026-1394/51/1A/06014
undefined