Abstract
A direct comparison of the standards for air kerma of the VSL, Netherlands, and of the Bureau International des Poids et Mesures (BIPM) was carried out in the 137Cs radiation beam of the BIPM in April 2016. The comparison result, evaluated as a ratio of the VSL and the BIPM standards for air kerma, is 0.9952 with a combined standard uncertainty of 3.8 × 10-3. The results for an indirect comparison made at the same time are consistent with the direct results at the level of 1 part in 104. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.
Main text
To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).