Key comparison BIPM.RI(I)-K9 of the absorbed dose to water standards of the LNE-LNHB, France, and the BIPM in medium-energy x-rays

, and

©2020 BIPM & IOP Publishing Ltd
, , Citation D T Burns et al 2020 Metrologia 57 06008DOI 10.1088/0026-1394/57/1A/06008

0026-1394/57/1A/06008

Abstract

A key comparison has been made between the absorbed dose to water standards of the LNE-LNHB, France, and the BIPM in the medium-energy x-ray range. The results show the standards to be in general agreement at the level of the expanded (k = 2) uncertainty of the comparison of 2 parts in 102. The results are presented in terms of degrees of equivalence for entry in the BIPM key comparison database.

Main text

To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).