Perturbation correction of a cylindrical thimble-type chamber in a graphite phantom for 60Co gamma rays

, and

Published under licence by IOP Publishing Ltd
, , Citation C E de Almeida et al 1989 Phys. Med. Biol. 34 1443 DOI 10.1088/0031-9155/34/10/007

0031-9155/34/10/1443

Abstract

An experimental determination has been made of the perturbation correction to be applied to a graphite cylindrical (thimble-type) ionisation chamber used for the measurement of absorbed dose in graphite. The results show that the magnitude of this correction increases with depth and depends on the size of the cavity. The knowledge of this correction makes possible the use of such a chamber as a standard dose in graphite for 60Co gamma rays.

Export citation and abstract BibTeX RIS

10.1088/0031-9155/34/10/007