Table of contents

Volume 44

Number 2, 1 February 1999

Previous issue Next issue

PAPERS

323

, , and

Several groups are developing ultra-miniature x-ray machines for clinical use in radiation therapy. Current systems are for interstitial radiosurgery and for intravascular insertion for irradiation to prevent re-stenosis. Typical generating voltages are low, in the 20 to 40 kV range. It is well established that the biological effectiveness of such low-energy photons is large compared with higher-energy gamma rays, because of the dominance of photoelectric absorption at low energies. We have used microdosimetric analyses to estimate RBEs for such devices, both at low doses and clinically relevant doses, relative to radiations from , , and Y. The RBEs at clinically relevant doses and dose rates for these low-energy x-ray sources are considerably above unity, both relative to and to photons, and also relative to and Y brachytherapy sources. As a function of depth, the overall effect of the change in dose and the change in beam spectrum results in beams whose biologically weighted dose (doseRBE) decreases with depth somewhat more slowly than does the physical dose. The estimated clinically relevant RBEs are sufficiently large that they should be taken into account during the treatment design stage.

335

, and

This paper describes the finite-difference time-domain (FDTD) analysis of antenna-body interaction effects occurring when chest-mounted 418 MHz radio transmitters are used for medical telemetry applications. Whole-body software models (homogeneous, layered and tissue-segmented) were developed for an adult male subject. Using an electrically small (300 ) planar loop antenna, calculated radiation efficiencies ranged between 33.5% and 39.2% for a whole-body model, and between 60.7% and 66.1% for a torso; radiation patterns were found to be largely independent of model composition. The computed radiation efficiency for a 21.5 kg phantom representing a six-year-old female was within 1.1 dB of measured results (actual body mass 28 kg) and well-correlated azimuthal radiation patterns were noted.

347

, , and

Membrane solubilization and osmotic fragility of rat erythrocytes irradiated in vivo with fast neutron fluences ranging from to using a source were measured instantaneously using a light scattering technique. The solubilization of erythrocyte membrane by a non-ionic detergent, octylglucoside (OG), was found to exhibit a two stage transition from vesicular form to mixed micellar form in the range of detergent concentrations 1.5-7.8 mM. The coexistence phase, vesicular/mixed micellar, was shifted towards higher detergent concentrations with increase in the neutron fluence, indicating increasing membrane resistance to the detergent and hence change in the natural membrane permeation properties. The technique shows an adequate sensitivity in detecting membrane damage in erythrocytes and has potential as a biophysical marker of radiation exposure. The osmotic fragility of irradiated erythrocytes shows a decreasing trend with increasing irradiation fluence measured directly and two weeks post-irradiation. Blood films photographed two weeks post-irradiation show developed elliptocytosis and crenated cell anaemia.

357

, , , and

The dose rate at point P at 0.25 cm in water from the transverse bisector of a straight catheter with an active stepping source (Nucletron microSelectron HDR source) with a dwell length of 2 cm was calculated using Monte Carlo code MCNP 4.A. The source step sizes were 1 cm and 0.25 cm. The Monte Carlo (MC) results were used for comparison with the results calculated with the Nucletron brachytherapy planning system (BPS) formalism, first with BPS variants and then with its respective MC calculated radial dose function and anisotropy function. The dose differences at point P calculated using the BPS formalism and variants are +15.4% and +3.1% for the source step size of 1 cm and 0.25 cm respectively. This reduction in dose difference is caused by the increased importance of errors in the anisotropy function with the smaller step size, which counter the errors in the radial dose function. Using the MC calculated radial dose function and anisotropy function with the BPS formalism, 1% dose calculation accuracy can be achieved, even in the near field, with negligible extra demand on computation time.

365

, , , , , , and

A coherent system for the use of scatter correction factors, determined at 10 cm depth, is described for dose calculations on the central axis of arbitrarily shaped photon beams. The system is suitable for application in both the fixed source-surface distance (SSD) and in the isocentric treatment set-up. This is in contrast to some other proposals where only one of these approaches forms the basis of the calculation system or where distinct quantities and data sets are needed. In order to derive the relations in the formalism, we introduced a separation of the phenomena related to the energy fluence in air and to the phantom scatter contribution to the dose. Both are used relative to quantities defined for the reference irradiation set-up. It is shown that dose calculations can be performed with only one set of basic beam data, obtained at a reference depth of 10 cm. These data consist for each photon beam quality of measured collimator and phantom scatter correction factors, in combination with a set of (percentage/relative) depth-dose or tissue-phantom ratio values measured along the central axis of the beam. Problems related to measurements performed at the depth of maximum absorbed dose, due to the electron contamination of the beam, are avoided in this way. Collimator scatter correction factors are obtained by using a mini-phantom, while phantom scatter correction factors are derived from measurements in a full scatter phantom in combination with the results of the mini-phantom measurements. For practical reasons the fixed SSD system was chosen to determine the data. Then, dose calculations in a fixed SSD treatment set-up itself are straightforward. Application in the isocentric treatment set-up needs simple conversion steps, while the inverse approach, from isocentric to fixed SSD, is described as well. Differences between the two approaches are discussed and the equations for the conversions are given.

383

, and

A methodology for the constrained customization of non-coplanar beam orientations in radiotherapy treatment planning has been developed and tested on a cohort of five patients with tumours of the brain. The methodology employed a combination of single and multibeam cost functions to produce customized beam orientations. The single-beam cost function was used to reduce the search space for the multibeam cost function, which was minimized using a fast simulated annealing algorithm. The scheme aims to produce well-spaced, customized beam orientations for each patient that produce low dose to organs at risk (OARs). The customized plans were compared with standard plans containing the number and orientation of beams chosen by a human planner. The beam orientation constraint-customized plans employed the same number of treatment beams as the standard plan but with beam orientations chosen by the constrained-customization scheme. Improvements from beam orientation constraint-customization were studied in isolation by customizing the beam weights of both plans using a dose-based downhill simplex algorithm. The results show that beam orientation constraint-customization reduced the maximum dose to the orbits by an average of 18.8 (, 1SD)% and to the optic nerves by 11.4 (, 1SD)% with no degradation of the planning target volume (PTV) dose distribution. The mean doses, averaged over the patient cohort, were reduced by 4.2 (, 1SD)% and 12.4 ( 1SD)% for the orbits and optic nerves respectively. In conclusion, the beam orientation constraint-customization can reduce the dose to OARs, for few-beam treatment plans, when compared with standard treatment plans developed by a human planner.

401

, , and

For renal dynamic studies, the COST B2 hybrid phantom is an example of an artificially created software phantom. Although this phantom is useful, it is not possible to implement the phantom in a self-consistent fashion to produce, for example, a collection of tracer in the bladder which is related to the flow from the kidneys. In this study control systems are used to provide a self-consistent model.

A feed-forward control system was designed for the transport of DTPA in the human body using SIMULINK. The system is based on a three-compartment model described by a set of differential equations with flow rates which may be set by the operator. The differential uptake in the kidneys may also be specified, while the flow of tracer through the renal parenchyma and collecting system of each kidney is determined using two-parameter retention functions.

Curves corresponding to normal or pathological conditions may be simulated for plasma, parenchyma, collecting system and bladder by appropriate selection of parameters. The system is user-friendly and can be used to simulate almost all conditions seen in patient studies. The next stage of using this information to design dynamic image simulations is in progress.

413

, and

We present a numerical model used to analyse the anisotropic electrical properties of frog muscle, measured in vivo. The model represents the anisotropic, irregularly shaped muscle as a set of cubic elements. We develop a finite difference method to calculate the electrical resistance between two electrodes inserted longitudinally or transversely into the muscle in terms of longitudinal and transverse muscle conductivities. Comparison of the measured impedance values with the calculated resistances yields the separate variation with frequency of the two conductivity components. We also compare the results of the numerical, finite difference method with those of two simple, analytical models.

423

, and

The spherical head model has been used in magnetoencephalography (MEG) as a simple forward model for calculating the external magnetic fields resulting from neural activity. For more realistic head shapes, the boundary element method (BEM) or similar numerical methods are used, but at greatly increased computational cost. We introduce a sensor-weighted overlapping-sphere (OS) head model for rapid calculation of more realistic head shapes. The volume currents associated with primary neural activity are used to fit spherical head models for each individual MEG sensor such that the head is more realistically modelled as a set of overlapping spheres, rather than a single sphere. To assist in the evaluation of this OS model with BEM and other head models, we also introduce a novel comparison technique that is based on a generalized eigenvalue decomposition and accounts for the presence of noise in the MEG data. With this technique we can examine the worst possible errors for thousands of dipole locations in a realistic brain volume. We test the traditional single-sphere model, three-shell and single-shell BEM, and the new OS model. The results show that the OS model has accuracy similar to the BEM but is orders of magnitude faster to compute.

441

and

The dose distributions in a spherical phantom irradiated by a Leksell Gamma Unit have been calculated using the EGS4 Monte Carlo code. In the simulation, the photon beams are considered to emanate from point sources, ignoring the scattering from source and collimating systems. The calculated results are in good agreement with results obtained with semiconductor diodes.

447

A mathematical model to calculate the theoretical cell survival probability (nominally, the cell survival fraction) is developed to evaluate preclinical treatment conditions for boron neutron capture therapy (BNCT). A treatment condition is characterized by the neutron beam spectra, single or bilateral exposure, and the choice of boron carrier drug (boronophenylalanine (BPA) or boron sulfhydryl hydride (BSH)). The cell survival probability defined from Poisson statistics is expressed with the cell-killing yield, the (n, ) reaction density, and the tolerable neutron fluence. The radiation transport calculation from the neutron source to tumours is carried out using Monte Carlo methods: (i) reactor-based BNCT facility modelling to yield the neutron beam library at an irradiation port; (ii) dosimetry to limit the neutron fluence below a tolerance dose (10.5 Gy-Eq); (iii) calculation of the (n, ) reaction density in tumours. A shallow surface tumour could be effectively treated by single exposure producing an average cell survival probability of - for probable ranges of the cell-killing yield for the two drugs, while a deep tumour will require bilateral exposure to achieve comparable cell kills at depth. With very pure epithermal beams eliminating thermal, low epithermal and fast neutrons, the cell survival can be decreased by factors of 2-10 compared with the unmodified neutron spectrum. A dominant effect of cell-killing yield on tumour cell survival demonstrates the importance of choice of boron carrier drug. However, these calculations do not indicate an unambiguous preference for one drug, due to the large overlap of tumour cell survival in the probable ranges of the cell-killing yield for the two drugs. The cell survival value averaged over a bulky tumour volume is used to predict the overall BNCT therapeutic efficacy, using a simple model of tumour control probability (TCP).

463

, , , , , and

A commercial three-dimensional (3D) inverse treatment planning system, Corvus (Nomos Corporation, Sewickley, PA), was recently made available. This paper reports our preliminary results and experience with commissioning this system for clinical implementation. This system uses a simulated annealing inverse planning algorithm to calculate intensity-modulated fields. The intensity-modulated fields are divided into beam profiles that can be delivered by means of a sequence of leaf settings by a multileaf collimator (MLC). The treatments are delivered using a computer-controlled MLC. To test the dose calculation algorithm used by the Corvus software, the dose distributions for single rectangularly shaped fields were compared with water phantom scan data. The dose distributions predicted to be delivered by multiple fields were measured using an ion chamber that could be positioned in a rotatable cylindrical water phantom. Integrated charge collected by the ion chamber was used to check the absolute dose of single- and multifield intensity modulated treatments at various spatial points. The measured and predicted doses were found to agree to within 4% at all measurement points. Another set of measurements used a cubic polystyrene phantom with radiographic film to record the radiation dose distribution. The films were calibrated and scanned to yield two-dimensional isodose distributions. Finally, a beam imaging system (BIS) was used to measure the intensity-modulated x-ray beam patterns in the beam's-eye view. The BIS-measured images were then compared with a theoretical calculation based on the MLC leaf sequence files to verify that the treatment would be executed accurately and without machine faults. Excellent correlation (correlation coefficients ) was found for all cases. Treatment plans generated using intensity-modulated beams appear to be suitable for treatment of irregularly shaped tumours adjacent to critical structures. The results indicated that the system has potential for clinical radiation treatment planning and delivery and may in the future reduce treatment complexity.

479

and

The implementation of biological optimization of radiation treatment plans is impeded by both computational and modelling problems. We derive an objective function from basic model assumptions which includes the normal tissue constraints as interior penalty functions. For organs that are composed of parallel subunits, a mean response model is proposed which leads to constraints similar to dose-volume constraints. This objective function is convex in the case when no parallel organs lie in the treatment volume. Otherwise, an argument is given to show that a number of local minima may exist which are near degenerate to the global minimum. Thus, together with the measure quality of the objective function, highly efficient gradient algorithms can be used. The number of essential biological model parameters could be reduced to a minimum. However, if the optimization constraints are given as TCP/NTCP values, Lagrange multiplier updates have to be performed by invoking comprehensive biological models.

495

, , , , , , , , , et al

An algorithm has been developed and experimentally verified for tomographic registration - a patient positioning method using internal anatomy and standard external fiducial marks. This algorithm improves patient set-up and verification to an accuracy sufficient for tomotherapy. By implementation of this technique, the time-consuming reconstruction process is avoided. Instead, offsets in the x, y and z directions are determined directly from sinogram data by an algorithm that utilizes cross-correlations and Fourier transforms. To verify the efficiency and stability of the algorithm, data were collected on the University of Wisconsin's dedicated tomotherapy research workbench. The experiment indicates offset statistical errors of less than mm for offsets up to 30 mm. With standard clinical techniques, initial patient offsets are expected to be less than 5 mm, so the 30 mm limitation is of no consequence. The angular resolution for the direction of patient translation is within the needed for tomotherapy.

509

, and

The purpose of this work was to establish procedures for the implementation of the Varian Enhanced Dynamic Wedge into a treatment planning system (TPS), based as much as possible on simple theoretical considerations and already available data.

A method is presented for the calculation (rather than measurement) of off-axis relative wedge transmission curves that are required by the TPS for relative dose calculations. We also present a method for absolute dose (monitor unit) calculations, based on the calculation of an effective wedge factor on the prescription point. A simple formula has been derived for the calculation of the effective wedge factor for the most general case, i.e. an arbitrary effective wedge angle, field size and prescription point.

Relative dose calculations have been verified by measurements performed on a Varian Clinac 2300C/D linear accelerator, for 6 MV and 20 MV photon energies. Monitor unit calculations have also been verified experimentally for several cases such as symmetric and asymmetric fields with prescription on the collimator axis or on the geometrical centre of the asymmetric field. The presented technique provides results within 2% for both relative and absolute dose calculations for clinically relevant cases.

525

, , , , , and

In order to evaluate the ability of the arthroscopic indentation instrument, originally developed for the measurement of cartilage stiffness during arthroscopy, to detect cartilage degeneration, we compared changes in the stiffness with the structural and constitutional alterations induced by enzymes on the tissue in vitro. The culturing of osteochondral plugs on Petri dishes was initiated in Minimum Essential Medium with Earle's salts and the baseline stiffness was measured. Then, the experimental specimens were digested using trypsin for 24 h, chondroitinase ABC or purified collagenase (type VII) for 24 h or 48 h ( n = 8-15 per group). The control specimens were incubated in the medium. After the enzyme digestion, the end-point stiffness was measured and the specimens for the microscopic analyses were processed. The proteoglycan (PG) distribution was analysed using quantitative microspectrophotometry and the quantitative evaluation of the collagen network was made using a computer-based polarized light microscopy analysis. Decrease of cartilage stiffness was found after 24 h trypsin (36%) and 48 h chondroitinase ABC (24%) digestion corresponding to a decrease of up to 80% and up to 30% in the PG content respectively. Decrease of the superficial zone collagen content or arrangement (78%, ) after 48 h collagenase digestion also induced a decrease (30%, ) in cartilage stiffness. We conclude that our instrument is capable of detecting early structural and compositional changes related to cartilage degeneration.

537

, and

A Monte Carlo computational model of a fluoroscopic imaging chain was used for deriving optimal technique factors for paediatric fluoroscopy. The optimal technique was defined as the one that minimizes the absorbed dose (or dose rate) in the patient with a constraint of constant image quality. Image quality was assessed for the task of detecting a detail in the image of a patient-simulating phantom, and was expressed in terms of the ideal observer's signal-to-noise ratio (SNR) for static images and in terms of the accumulating rate of the square of SNR for dynamic imaging. The entrance air kerma (or air kerma rate) and the mean absorbed dose (or dose rate) in the phantom quantified radiation detriment. The calculations were made for homogeneous phantoms simulating newborn, 3-, 10- and 15-year-old patients, barium and iodine contrast material details, several x-ray spectra, and for imaging with or without an antiscatter grid. The image receptor was modelled as a CsI x-ray image intensifier (XRII). For the task of detecting low- or moderate-contrast iodine details, the optimal spectrum can be obtained by using an x-ray tube potential near 50 kV and filtering the x-ray beam heavily. The optimal tube potential is near 60 kV for low- or moderate-contrast barium details, and 80-100 kV for high-contrast details. The low-potential spectra above require a high tube load, but this should be acceptable in paediatric fluoroscopy. A reasonable choice of filtration is the use of an additional 0.25 mm Cu, or a suitable K-edge filter. No increase in the optimal tube potential was found as phantom thickness increased. With the constraint of constant low-contrast detail detectability, the mean absorbed doses obtained with the above spectra are approximately 50% lower than those obtained with the reference conditions of 70 kV and 2.7 mm Al filter. For the smallest patient and x-ray field size, not using a grid was slightly more dose-efficient than using a grid, but when the patient size and field size were increased a fibre interspaced grid resulted in lower doses than imaging without a grid. For a 15-year-old patient the mean absorbed doses were up to 40% lower with this grid than without the grid.

561

, and

In this paper, we present reconstruction results from helical cone-beam CT data, obtained using a simple and fast algorithm, which we call the CB-SSRB algorithm. This algorithm combines the single-slice rebinning method of PET imaging with the weighting schemes of spiral CT algorithms. The reconstruction is approximate but can be performed using 2D multislice fan-beam filtered backprojection. The quality of the results is surprisingly good, and far exceeds what one might expect, even when the pitch of the helix is large. In particular, with this algorithm comparable quality is obtained using helical cone-beam data with a normalized pitch of 10 to that obtained using standard spiral CT reconstruction with a normalized pitch of 2.

571

and

Normalization in positron emission tomography (PET) is the process of ensuring that all lines of response joining detectors in coincidence have the same effective sensitivity. In three-dimensional (3D) PET, normalization is complicated by the presence of a large proportion of scattered coincidences, and by the fact that cameras operating in 3D mode encounter a very wide range of count-rates. In this work a component-based normalization model is presented which separates the normalization of true and scattered coincidences and accounts for variations in normalization effects with count-rate. The effects of the individual components in the model on reconstructed images are investigated, and it is shown that only a subset of these components has a significant effect on reconstructed image quality.

595

, , , , and

The thermoluminescent dosimeter (TLD) method has been proposed as a useful tool for estimating internal radiation absorbed dose in nuclear medicine. An efficient approach to verify the accuracy of the TLD method has been performed in this study. Under the standard protocol for 2-[F-18]fluoro-2-deoxy-D-glucose , whole body PET experiments and simultaneous body surface dose measurements by TLDs were performed on six normal volunteers. By using the body surface dose measured with TLDs, the cumulated activities of nine source organs were estimated with a mathematical unfolding technique for three different initial guesses. The accuracy of the results obtained by the TLD method was investigated by comparison with the actual cumulated activity of the same source organs measured by whole body PET. The cumulated activities of the source organs obtained by the TLD method and whole body PET show a significant correlation (correlation coefficient, , level of confidence, ) with each other. The mean effective doses in this study are obtained from the TLD method and obtained from the whole body PET. Good agreement between the results of the TLD method and whole body PET was observed.

607

, , , , , and

Non-invasive detection of small temperature changes (C) is pivotal to the further advance of regional hyperthermia as a treatment modality for deep-seated tumours. Magnetic resonance (MR) thermography methods are considered to be a promising approach. Four methods exploiting temperature-dependent parameters were evaluated in phantom experiments. The investigated temperature indicators were spin-lattice relaxation time , diffusion coefficient D, shift of water proton resonance frequency (water PRF) and resonance frequency shift of the methoxy group of the praseodymium complex (Pr probe). The respective pulse sequences employed to detect temperature-dependent signal changes were the multiple readout single inversion recovery (T One by Multiple Read Out Pulses; TOMROP), the pulsed gradient spin echo (PGSE), the fast low-angle shot (FLASH) with phase difference reconstruction, and the classical chemical shift imaging (CSI). Applying these sequences, experiments were performed in two separate and consecutive steps. In the first step, calibration curves were recorded for all four methods. In the second step, applying these calibration data, maps of temperature changes were generated and verified. With the equal total acquisition time of approximately 4 min for all four methods, the uncertainties of temperature changes derived from the calibration curves were less than C (Pr probe C, water PRF C, C and C). The corresponding maps of temperature changes exhibited slightly higher errors but still in the range or less than C (C, C, C, C respectively). The calibration results indicate the Pr probe method to be most sensitive and accurate. However, this advantage could only be partially transferred to the thermographic maps because of the coarse matrix of the classical CSI sequence. Therefore, at present the water PRF method appears to be most suitable for MR monitoring of small temperature changes during hyperthermia treatment.

NOTES

N9

, and

Two methods are compared for calculating the field-size dependence of the phantom scatter component of dose for x-ray beams. One model sums three Gaussian distributions; the other model is a two-parameter function. With a measurement of the beam quality as input to determine parameters, both models accurately reproduce the relative phantom scatter. However, there are important differences between the models. For all beam energies, the two-parameter model characterizes the absolute phantom scatter as a function of depth and field size, while, also for all beam energies, the six-parameter Gaussian model characterizes the relative phantom scatter at a single depth of 10 cm. For small field sizes, the phantom scatter calculated from the two-parameter model agrees with Monte Carlo calculations better than the Gaussian model. In the Gaussian model, the parameters can be obtained for beam energies between and 25 MV by linear interpolation based on the measured beam quality. In the two-parameter model, and for energies above 4 MV, the parameters can be obtained using linear functions of the dose-weighted average linear attenuation coefficient, which is related to beam quality.

N15

, and

A technique is described for manually outlining a volume of interest (VOI) in a three-dimensional SPECT dataset. Regions of interest (ROIs) are drawn on three orthogonal maximum intensity projections. Image masks based on these ROIs are backprojected through the image volume and the resultant 3D dataset is segmented to produce the VOI. The technique has been successfully applied in the exclusion of unwanted areas of activity adjacent to the brain when segmenting the organ in SPECT imaging using HMPAO. An example of its use for segmentation in tumour imaging is also presented. The technique is of value for applications involving semi-automatic VOI definition in SPECT.

LETTER TO THE EDITOR

001

and

X-ray diagnostics gives the largest contribution to the population dose from man-made radiation sources. Strategies for reduction of patient doses without loss of diagnostic accuracy are therefore of great interest to society and have been focussed in general terms by the ICRP (ICRP 1996) through the introduction of the concept of diagnostic reference levels. The European Union has stimulated research in the field, and, based on patient dose measurements and radiologists' appreciation of acceptable image quality, good radiographic techniques have been identified and recommended (EUR 1996a, b) for conventional screen-film imaging. These efforts have resulted in notable dose reductions in clinical practices (Hart et al 1996).

In spite of 100 years of use of x-rays for diagnostics, the choice of technique parameters still relies to a great extent on experience. Scientific efforts to optimize the choice in terms of finding the parameter settings which yield sufficient image quality at the lowest possible cost in dose are still rare. True optimization requires (1) estimation of the image quality needed to make a correct diagnosis and (2) methods to investigate all possible means of achieving this image quality in order to be able to decide which of them gives the lowest dose.

The paper by Tapiovaara, Sandborg and Dance published in this issue of Physics in Medicine and Biology (pages 537-559) addresses the optimization of paediatric fluoroscopy, a timely and important topic. Fluoroscopy procedures, used to guide x-ray examinations or interventional procedures, are little standardized and may result in high dose levels; radiation exposure in childhood is likely to result in a higher lifetime risk than the same exposure later in life. The authors represent an interesting mix of expertise within various scientific fields: the theory of medical imaging and assessment of image quality, the physics of diagnostic radiology and radiation dosimetry. They provide good insights into the technical performance and limitations of clinical x-ray equipment and the functioning of x-ray image intensifiers (XRIIs), and discuss the problem in an educational manner. The authors have succeeded in producing a paper that combines high scientific merit and valuable practical guidelines towards the optimization of paediatric fluoroscopy and radiography. This information, if properly utilized by practitioners, will contribute to significant (50%) reduction in radiation doses without sacrificing image quality and diagnostic accuracy.

Fluoroscopy using XRIIs is one of the x-ray procedures that allows the possibility of bringing patient doses to an absolute minimum: quantum-noise due to the inevitable stochastic nature of the interactions of x-rays with the image receptor will ultimately limit the possibility for further dose reduction. In a well designed (quantum-noise limited) system, patient dose ( D) increases proportionally with the square of signal-to-noise ratio (SNR). The SNR of the ideal observer (ICRU 1996) may be used as image quality descriptor. To ascertain minimum patient dose, the just detectable SNR level of critical image details in a given diagnostic procedure should be determined. The dose-to-information conversion factor SNR2/D is independent of patient dose, thus providing a useful figure-of-merit for optimization of the technique parameters of an imaging task. In the cited study, the strategy for optimization is to maximize the SNR2/D ratio, leaving the absolute requirement on SNR (and patient dose) to be determined by the user. Technique factors which strongly influence the SNR2/D ratio are the energy spectrum (tube potential and total filtration) and the choice of anti-scatter technique. These can be adjusted by the user of clinical x-ray equipment and are focussed in the paper. A powerful tool used in executing the study is a carefully developed and validated computational model of the imaging chain (Tapiovaara and Sandborg 1995) allowing simultaneous calculation of SNR and D, and flexible variation of the parameters. The results indicate that optimal parameter settings may differ significantly from those recommended on the basis of previous experience: low tube potential techniques are suggested to improve dose efficiency as well as use of appropriately designed anti-scatter (fibre) grids. The detection task, or the type of important detail, also influences the choice of preferred technique.

The current study has important implications for other x-ray imaging procedures which, like XRII fluoroscopy, are not restricted by the requirement of optical density as in conventional screen-film imaging. It suggests the importance of optimizing the imaging techniques for digital radiographic systems so that their advantages over conventional systems can be fully utilized. Another natural extension of the current work is high dose fluoroscopy in interventional procedures which has been a major concern of radiation risks to patients (Shope 1996). A search for improved techniques to minimize exposure in general fluoroscopy is a continued challenge to imaging physicists. An even greater challenge is posed to the fluoroscopic x-ray equipment manufacturers. Some systems will need to be redesigned to allow users greater flexibility of selecting a desired strategy for the automatic brightness compensation control. It will also be essential to provide higher output for fluoroscopic systems to accommodate low kV and heavy filtration operations.

In summary, the paper by Tapiovaara et al (1999) is an important contribution to development of dose efficient techniques which will influence future practices and manufacturers of x-ray equipment. Scientific investigations such as this one will serve as a driving force for practical implementation of these dose efficient techniques.

References

EUR 1996a European guidelines on quality criteria for diagnostic radiographic images EUR 16260 EN (Luxembourg: Office of Official Publications of the European Communities)

EUR 1996b European guidelines on quality criteria for diagnostic radiographic images in paediatrics EUR 16261 EN (Luxembourg: Office of Official Publications of the European Communities)

Hart D, Hillier M C, Wall B F, Shrimpton P C and Bungay D 1996 Doses to patients from medical x-ray examinations in the UK - 1995 Review NRPB-R289 (London: HMSO)

ICRP 1996 Radiological protection and safety in medicine ( ICRP Publication 73) Ann. ICRP 26 (Oxford: Pergamon)

ICRU 1996 Medical imaging - the assessment of image quality ICRU Report 54 (Bethesda, MD: International Commission on Radiation Units and Measurements)

Shope T B 1996 Radiation-induced skin injuries from fluoroscopy RadioGraphics 16 1195-9

Tapiovaara M J and Sandborg M 1995 Evaluation of image quality in fluoroscopy by measurements and Monte Carlo calculations Phys. Med. Biol. 40 589-607

Tapiovaara M J, Sandborg M, and Dance D R 1999 A search for improved technique factors in paediatric fluoroscopy Phys. Med. Biol. 44 537-59