Introducing a system for automated control of rotation axes, collimator and laser adjustment for a medical linear accelerator

, , and

Published 10 April 2003 Published under licence by IOP Publishing Ltd
, , Citation Peter Winkler et al 2003 Phys. Med. Biol. 48 1123 DOI 10.1088/0031-9155/48/9/303

0031-9155/48/9/1123

Abstract

Mechanical stability and precise adjustment of rotation axes, collimator and room lasers are essential for the success of radiotherapy and particularly stereotactic radiosurgery with a linear accelerator. Quality assurance procedures, at present mainly based on visual tests and radiographic film evaluations, should desirably be little time consuming and highly accurate. We present a method based on segmentation and analysis of digital images acquired with an electronic portal imaging device (EPID) that meets these objectives. The method can be employed for routine quality assurance with a square field formed by the built-in collimator jaws as well as with a circular field using an external drill hole collimator. A number of tests, performed to evaluate accuracy and reproducibility of the algorithm, yielded very satisfying results. Studies performed over a period of 18 months prove the applicability of the inspected accelerator for stereotactic radiosurgery.

Export citation and abstract BibTeX RIS