Dual-energy imaging in full-field digital mammography: a phantom study

, , , , , , and

Published 17 June 2003 Published under licence by IOP Publishing Ltd
, , Citation A Taibi et al 2003 Phys. Med. Biol. 48 1945 DOI 10.1088/0031-9155/48/13/307

0031-9155/48/13/1945

Abstract

A dual-energy technique which employs the basis decomposition method is being investigated for application to digital mammography. A three-component phantom, made up of plexiglas, polyethylene and water, was doubly exposed with the full-field digital mammography system manufactured by General Electric. The 'low' and 'high' energy images were recorded with a Mo/Mo anode-filter combination and a Rh/Rh combination, respectively. The total dose was kept within the acceptable levels of conventional mammography. The first hybrid images obtained with the dual-energy algorithm are presented in comparison with a conventional radiograph of the phantom. Image-quality characteristics at contrast cancellation angles between plexiglas and water are discussed. Preliminary results show that a combination of a standard Mo-anode 28 kV radiograph with a Rh-anode 49 kV radiograph provides the best compromise between image-quality and dose in the hybrid image.