It is shown that, in the design of magnetic probes for plasma field measurements, attention must be given to the space resolving power of the coils. A technique is described, which, by mechanical means, gives an exact alignment of the individual coils in a multiple probe without any need for further manual adjustment. Integration of the signals by means of transistorized Miller integrators, placed at the probe end of the transmitting cables, gives a good signal-to-noise ratio. The described system has an overall sensitivity of 1 mv G-1 at an RC time of 8 msec, and the bandwidth extends to over 2 Mc/s.