Calibration of a two-phase xenon time projection chamber with a 37Ar source

, , , , , , , , , and

Published 4 August 2017 © 2017 IOP Publishing Ltd and Sissa Medialab
, , Citation E.M. Boulton et al 2017 JINST 12 P08004 DOI 10.1088/1748-0221/12/08/P08004

1748-0221/12/08/P08004

Abstract

We calibrate a two-phase xenon detector at 0.27 keV in the charge channel and at 2.8 keV in both the light and charge channels using a 37Ar source that is directly released into the detector. We map the light and charge yields as a function of electric drift field. For the 2.8 keV peak, we calculate the Thomas-Imel box parameter for recombination and determine its dependence on drift field. For the same peak, we achieve an energy resolution, Eσ/Emean, between 9.8% and 10.8% for 0.1 kV/cm to 2 kV/cm electric drift fields.

Export citation and abstract BibTeX RIS