We evaluated a compact and cost-effective radio thin-layer chromatography (radio-TLC) scanner for the quality control (QC) of radiopharmaceuticals. We adapted a scintillation detector, which is a Gd3Al2Ga3O12 (GAGG:Ce) scintillation crystal array coupled with a photodiode array. The performance of the scintillator array-based radio-TLC was compared with that of a commercial device. We scanned 1 μCi/μL of Tc-99m and F-18 with each device. The difference between the ROI count ratios of the developed and commercial scanners was less than 1.2%. Our scanner is sensitive enough to take measurements for a radiochemical purity test.

Export citation and abstract BibTeX RIS

Published by IOP Publishing Ltd on behalf of Sissa Medialab. Original content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.