A GEANT4 Monte Carlo simulation to describe the time response of a coupled SiPM and LYSO detection system

, , , and

Published 13 June 2014 © 2014 IOP Publishing Ltd and Sissa Medialab srl
, , Citation E Leming et al 2014 JINST 9 C06008 DOI 10.1088/1748-0221/9/06/C06008

1748-0221/9/06/C06008

Abstract

In recent years the silicon photomultiplier has been investigated as an alternative to the traditional photomultiplier tube in a range of applications, including Time-of-flight Positron Emission Tomography (TOF-PET). In this paper we discuss a GEANT4 simulation framework, which has been developed to drive the design of a scalable TOF-PET apparatus to be built at the Rutherford Appleton Laboratory, UK. First results presented in this paper simulate the response of an Hamamatsu Multi-Pixel Photon Counter (S10362-33-050c) coupled to LYSO scintillating crystals, with focus on the timing response of coincidence signals.