Paper The following article is Open access

Ferrum nano particles and multiwall carbon nano tubes based electrode as FIA detector for determination of amino acids in hypothalamus microdialysis fluids

, , , , and

Published under licence by IOP Publishing Ltd
, , Citation L Sun et al 2017 IOP Conf. Ser.: Mater. Sci. Eng. 167 012025 DOI 10.1088/1757-899X/167/1/012025

1757-899X/167/1/012025

Abstract

An amperometric electrode based on multiwall carbon nanotubes (MWCNTs) and Fe nanoparticles (NPs) has been successfully fabricated. Combined with Flow Injection Analysis (FIA) and chromatography separation column, the electrode exhibits linear response in the concentration range of 0.1 -12 μM and the sensitivity of 30.0 nA μM−1 for most of amino acids. The determination of 17 amino acids in the hypothalamus microdialysis fluids of guinea pigs, illustrates that the electrode is a powerful tool to investigate physiology and pathology mechanisms

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1757-899X/167/1/012025